John Hay Biomedical · Principles of Biomedical Science

Find the first tap.
Follow it to silence.

Learn what to listen for when measuring blood pressure with a cuff and stethoscope. Watch a labeled pressure gauge, hear optional synthetic pulses, and connect each change to the pressure in a fictional artery.

Fictional case BP-A40-second modelOptional synthetic audio
01 · What to know

The sound comes from the cuffed artery.

Your heart creates a pressure pulse with each beat. Systolic pressure is the peak arterial pressure during a heartbeat. Diastolic pressure is the lower arterial pressure between beats. Blood pressure is usually reported as systolic over diastolic, in millimeters of mercury (mmHg).

Korotkoff sounds are the taps and other sounds heard over a compressed artery as a cuff slowly deflates. Cuff compression changes how pulse waves move the arterial wall and nearby tissue. These vibrations reach the stethoscope while blood passes through the partly compressed artery. These are different sounds from the heart-valve “lub-dub” heard on the chest. Sources 1, 2, and 4

Above systolic

The cuff pressure is high enough to block blood flow through the compressed artery. There are no Korotkoff taps.

Between the two pressures

As pressure falls, blood passes through the partly compressed artery during each beat. Recurring sounds begin. The onset of repeated taps, phase I, marks systolic pressure.

Near the lower boundary

Sounds can become muffled, called phase IV. Muffling is a change in sound, not the usual adult diastolic endpoint.

Below diastolic

Korotkoff sounds disappear, called phase V. This is the usual adult diastolic endpoint. Blood continues to flow; silence does not mean flow has stopped.

Figure BP-1. Three stages of a compressed artery during cuff deflationLeft: cuff pressure above systolic compresses the artery and blocks flow. Center: between systolic and diastolic, pulsed passage through the compressed artery produces recurring Korotkoff sounds. Right: below diastolic, the artery stays open and blood flows while Korotkoff sounds disappear. Simplified original teaching illustration, not a scale drawing.ABOVE SYSTOLICBETWEEN PRESSURESBELOW DIASTOLICCuff compresses arteryPulsed passageArtery stays openFlow blockedRecurring soundFlow continues silentlyNo Korotkoff soundWall and tissue motionNo Korotkoff soundduring pulsed flow
Figure BP-1 · What to notice: Cuff pressure changes the artery's behavior. Both the blocked and fully open states can be silent. This original simplified illustration shows the concept, not exact anatomy or the full sound mechanism.
02 · Directions and listening model

One pressure scale.
One beat at a time.

In fictional case BP-A, the model's arterial pressures are 118/76 mmHg. The cuff begins at 150 mmHg and falls by 2 mmHg each second. A modeled heartbeat occurs once per second. First, predict when the first tap will occur. Then play the model and compare the sound, text, and gauge.

This is a stylized teaching model. The synthesized pulses are not patient recordings and are not a clinically validated sound trainer. This page measures no blood pressure and uses no microphone. Your speakers, headphones, and volume setting do not calibrate a stethoscope or establish a physical sound level.

Sound is optional and starts only after you choose it. Begin with low device volume. Every pulse and silent beat also has a text and visual equivalent.

A digital control, not a measured dB level.

Audio is off. The complete visual model is available.

Figure BP-2. Simulated cuff-pressure gaugeSimulated pressure begins at 150 millimeters of mercury. Small ticks show 2 millimeters of mercury; numbered intervals are 20. The digital text gives the same modeled pressure.SIMULATED CUFFmmHgSmall tick = 2 mmHg
Figure BP-2 · The needle and digital readout share the same modeled pressure. This is a teaching drawing, not a working instrument.
Simulated cuff pressure
150 mmHg
0.0 / 40.0 seconds · 2 mmHg per second
Ready to begin

The cuff is above the model's systolic pressure. Flow is blocked through the compressed artery; no Korotkoff sound is expected.

No beat observed yet.

Moving the time control pauses playback. Arrow keys move by 0.1 second. “Jump before first tap” goes to 14 seconds; “Jump before last tap” goes to 34 seconds.

First-tap observation: not marked.

Disappearance observation: not marked.

Your marks record the current simulated gauge pressure. Reaction time and the interval between beats affect a click; this activity gives no clinical score. Use the evidence review to explain your observations. Reset clears these temporary marks.

Listen for a repeated event

A single tap can be noise. In a real manual measurement, verify the start with repeated beats and use the instructor's full procedure. In this model, taps at 118 and 116 mmHg establish the repeated onset. Sounds become softer near the end. The last modeled sound occurs at 78 mmHg; the next expected beat, at 76 mmHg, is silent.

The model sets a sharp boundary for learning. Real observations have finite beat spacing, variable sound quality, and measurement uncertainty. This model does not claim 1 mmHg clinical accuracy.

03 · Worked evidence review

Read the change.
Explain the endpoint.

Use the model's known values to connect what you heard or saw with a reported blood pressure. You can inspect the complete case with audio off.

Choose “Show the model's evidence” above to reveal the labeled timeline, exact event table, and worked report. The evidence can be opened before, during, or after playback.

04 · Troubleshooting and your work

Make the listening conditions work.

A more expensive stethoscope cannot correct every listening problem. Before repeating a supervised classroom measurement, check the setup with your teacher. A quiet, well-placed chestpiece matters.

  1. Fit and activate the stethoscope. Earpieces point forward toward the nose. If the model has a selectable chestpiece side, verify the intended side is active. Follow its manufacturer's directions.
  2. Locate the listening site. Place the chestpiece, the round listening end, lightly over the brachial artery just below the cuff, in the inner elbow region. Keep it out from under the cuff. Avoid pressing hard or rubbing tubing, clothing, or the chestpiece.
  3. Set the conditions. Use the correct cuff size for the arm. Support the arm with the cuff at heart level. Keep the listener, subject, and room quiet. See the illustrated guide for the full preparation sequence.
  4. Slow the release. Deflate at about 2 mmHg per second. Listen through the whole sequence: first repeated taps, changing sounds, then disappearance. Fast release can skip the useful events.
  5. Stop and ask. If there is discomfort, or you cannot hear the expected sounds, fully deflate the cuff and ask the teacher to check placement, equipment, and technique. Do not keep pumping higher to search for a sound.

Record and explain

Play BP-A with audio off. Use the visual and text signals to mark the first tap and sound disappearance. Confirm that the first tap is followed by another tap. Review the evidence table. Explain why the last audible beat at 78 is different from the programmed silent boundary at 76.

Reflect and revise

If you wish, repeat with the synthetic sound enabled. Describe the recurring tap and the muffled ending. State why hearing silence alone cannot distinguish an overinflated cuff from cuff pressure below diastolic.

Your next step: Use the Manual Blood Pressure Illustrated Guide for the complete supervised workflow and labeled stethoscope placement. This practice is optional, stores no responses, and creates no new graded submission. All facts needed for the prompts are on this page.
Quick vocabulary

Cuff pressure: pressure applied by the inflatable arm cuff. Brachial artery: the upper-arm artery used for this listening method. Korotkoff sounds: sounds from the compressed artery and surrounding tissues as the cuff deflates. Systolic: peak arterial pressure during a heartbeat. Diastolic: lower arterial pressure between beats. mmHg: millimeters of mercury, a pressure unit. Phase I: repeated sound onset. Phase IV: muffling. Phase V: sound disappearance.

Source notes

The evidence behind the model.

Clinical measurement sources support the procedure and endpoints. The 40-second timeline, 118/76 values, and synthetic sound timbre are original fictional teaching choices.

  1. American Heart Association scientific statement (2019) · Measurement of Blood Pressure in Humans. Preparation, cuff and stethoscope technique, slow deflation, and auscultatory endpoints.
  2. New England Journal of Medicine · Videos in Clinical Medicine: Blood Pressure Measurement. Korotkoff phases and measurement technique.
  3. CDC · Measuring Your Blood Pressure. Positioning and measurement preparation.
  4. Primary research (2023) · The fundamental mechanisms of the Korotkoff sounds generation. Arterial-wall and tissue motion underlying recorded Korotkoff sounds. The synthesized taps here are illustrative and do not reproduce that study's measurements.